Development of a slice-specific ct dose library with interpolation methods for organ dosimetry in MIRDct

In brief

CT dose interpolation estimates most organ doses within 5% of simulations

In computer simulations using selected adult female and pediatric male phantoms, interpolating dose estimates across 1-centimeter CT slices kept estimates within 5% of simulation results for most tissues; all errors were below 7%. The method could reduce the computing needed for organ-dose estimates and is planned for the MIRDct tool, but its performance in clinical use remains to be established.

Journal
Journal of applied clinical medical physics (Q2)
Published
1 December 2026
Study design
Unclassified
Evidence level
Level 5, Expert Opinion (CEBM 5)
Authors
Jared M Baggett, Laura E Dinwiddie, Robert J Dawson, Juan C Ocampo Ramos, Gunjan Kayal, Lukas M Carter, et al.
PMID
43059133
DOI
10.1002/acm2.70914

Why clinicians should know about it

  • Picked for Medical Physics (top studies of the week, 22 November 2026): Monte Carlo CT organ-dose library and interpolation validation

Abstract

OBJECTIVES: To develop a slice-specific dose library and supporting interpolation scheme for proper estimation of organ doses from CT scans based on the UF/MSK mesh-type adult and pediatric phantom series. METHODS: Monte Carlo (MC) simulations were performed using the Particle and Heavy Ion Transport code System (PHITS) for selected adult female and pediatric male phantoms across a user defined CT source term. Dose profiles were calculated on a 1-cm slice basis and categorized by tissue type (distributed vs. non-distributed). Linear interpolation of slice doses within a height series was evaluated using multiple phantom subsets and compared against MC results. RESULTS: Slice-by-slice linear interpolation produced total organ dose estimates within 5% of MC values for most studied tissues, with all errors below 7%. Mean errors for abdomen, chest-abdomen-pelvis and cardiac scan ranges were also below 5%, and mean errors for individual slices were below 3%. The method performed well across various interpolation schemes, especially when phantoms were selected based on similar construction methods. CONCLUSIONS: This approach enables accurate, scalable organ dose estimation with reduced computational demands. The resulting dose library and interpolation method will be integrated into the MIRDct platform, a dosimetry tool for calculating total organ dose values across a wide range of CT parameters. MIRDct is freely available to the medical imaging community at MIRDsoft.org.