Intraoperative Impedance Planimetry Reveals Functional Sleeve Geometry Beyond Fixed Bougie Calibration
In brief
Fixed bougie sizing leaves sleeve geometry variable; narrower sleeves linked to nausea
In this prospective pilot, measurements during sleeve gastrectomy found substantial variation despite use of a standard 32-French bougie. Smaller minimum diameter and cross-sectional area were associated with more postoperative nausea during the hospital stay, but the proposed cutoffs came from this cohort. Larger studies must confirm whether these measurements reliably predict outcomes before they guide surgery.
- Journal
- Obesity surgery (Q1)
- Published
- 18 November 2026
- Study design
- Prospective / inception cohort
- Evidence level
- Level 2, Moderate (CEBM 2b)
- Authors
- Michael de Cillia, Jonas Bolm, Martin Grünbart, Michael Weitzendorfer, Hannes Hoi, Alexandra Majerus, et al.
- PMID
- 43050038
- DOI
- 10.1007/s11695-026-09047-9
Why clinicians should know about it
- Picked for Bariatric and Metabolic Surgery (top studies of the week, 22 November 2026): Intraoperative sleeve geometry linked to postoperative nausea
Abstract
BACKGROUND: Rigid bougies are commonly used for intraoperative sleeve calibration in standardized sleeve gastrectomy to ensure a reproducible sleeve geometry. Current evidence on bougie size and its impact on postoperative outcome and sleeve shape remains inconclusive and inconsistent. This pilot study was conducted to assess whether a fixed bougie size leads to reproducible sleeve configuration and to evaluate the relationship between early postoperative outcome and objective intraoperative luminal sleeve geometry measurements. METHODS: A prospective analysis was performed in patients providing informed consent for sleeve gastrectomy (SG). SG using a 32Fr bougie and the intraoperative measurements followed a meticulous standardized protocol. Distensibility index (DI), cross-sectional area (CSA), minimal diameter (Dmin), high-pressure zone (HPZ) and intra-balloon pressure (IBP) served as objective measurement parameters of functional sleeve geometry, and correlation analyses with postoperative outcomes were performed. RESULTS: SG with objective intraoperative sleeve evaluation was technically feasible in all patients and no device-related adverse events were observed in this pilot cohort. Luminal measurements showed exploratory associations with early clinical outcome, with lower Dmin and CSA associated with a higher occurrence of postoperative nausea (Dmin: 95% CI -1.00 to -0.357; CSA: 95% CI -0.947 to -0.188], respectively). Exploratory ROC analysis identified cohort-derived candidate cut-off values of 18.15 mm for Dmin and 259.5 mm² for CSA. CONCLUSION: Despite standardized SG with a fixed bougie size, substantial variability persisted in luminal EndoFLIP™ parameters, indicating that bougie diameter alone does not fully determine the measured geometry of the newly constructed sleeve. Intraoperative impedance planimetry identified Dmin and CSA as exploratory candidate markers associated with postoperative nausea during index hospital stay. Further prospective studies are required to assess its value as a step towards physiology-guided SG.
Abstract as published, via PubMed.
For healthcare professionals. The summary is generated by AI from the published abstract, and the evidence level is assigned automatically from the study design on the Oxford CEBM hierarchy. Neither is medical advice. Read the full paper before changing practice.