Quantitative static contrast-enhanced phantom study to establish performance of advanced cone beam computed tomography for functional imaging applications
- Journal
- Physics and imaging in radiation oncology (Q1)
- Published
- 7 November 2026
- Study design
- Unclassified
- Evidence level
- Level 5, Expert Opinion (CEBM 5)
- Authors
- Riley McMaster, Raanan Marants, Hedi Mohseni, Catherine Coolens
- PMID
- 43032152
- DOI
- 10.1016/j.phro.2026.101125
Why clinicians should know about it
- Picked for Medical Physics (paper of the day, 19 November 2026): Phantom study measures quantitative CBCT image performance
Abstract
BACKGROUND AND PURPOSE: Dynamic contrast-enhanced computed tomography (DCE-CT) is a quantitative imaging technique that measures vascular perfusion. Extending this to cone beam CT (CBCT) would enable functional measurements on radiotherapy machines. An advanced CBCT imaging solution, with enhanced image quality and rapid acquisitions, offers potential. However, the static quantitative and spatial performance of advanced CBCT relevant to DCE imaging has not been characterized. MATERIALS AND METHODS: To address this, the image quality of advanced CBCT was compared with standard CBCT and CT using contrast-enhanced phantoms with DCE-specific characteristics not previously applied to CBCT. A contrast enhancement CT (CECT) phantom (acrylic cylinder with capsule inserts) and a humanoid phantom (human torso with a liver containing tumour-mimicking spheres) were imaged across different acquisition settings. RESULTS: Using the CECT phantom, advanced CBCT demonstrated improved CT number accuracy compared to standard CBCT (<23.9 vs. <78.9 HU) and greater CT number constancy (<8.4 vs. <12.9 HU). Iterative reconstruction on advanced CBCT offered improvements over filtered back-projection. Using the humanoid phantom, findings were confirmed with CT number accuracy (<7.1 vs. <42.8 HU), CT number variation (<11.3 vs. <49.8 HU) and volume accuracy (<128.8 vs. <1067.3 mm3). Advanced CBCT achieved spatial resolutions of 2.3-5.0 lp/cm across different acquisitions and measurements, meeting the 5 lp/cm clinical benchmark under certain conditions. It also displayed stable CT number-density calibration. CONCLUSIONS: Advanced CBCT compared to standard CBCT improved image quality, approaching CT-level performance. Findings demonstrated robust volume and contrast-specific quantification under DCE conditions, supporting further evaluation of functional CBCT.
Abstract as published, via PubMed.
For healthcare professionals. The summary is generated by AI from the published abstract, and the evidence level is assigned automatically from the study design on the Oxford CEBM hierarchy. Neither is medical advice. Read the full paper before changing practice.