Low-Dose PCCT-Based Liver Volume Perfusion CT with Phase-Specific Reconstruction Optimization for Morphologic

In brief

Photon-counting liver CT cuts radiation dose by 19% versus a simulated protocol

In 62 patients, a one-stop photon-counting CT protocol delivered an average effective dose of 25.9 mSv, compared with 31.9 mSv for a simulated four-phase CT protocol plus perfusion imaging. Most perfusion-derived images were rated diagnostically acceptable, but quality depended on reconstruction and some portal venous images fell short, so broader clinical validation is still needed.

Journal
Academic radiology (Q1)
Published
10 November 2026
Study design
Cohort / observational study
Evidence level
Level 3, Low (CEBM 3b)
Authors
Yong Chen, Weiping Ma, Bo Ma, Xiaobin Ma, Lu Chen, Yongwei Liu, et al.
PMID
43009363
DOI
10.1016/j.acra.2026.10.024

Why clinicians should know about it

  • Picked for Medical Physics (paper of the day, 11 November 2026): PCCT image quality and radiation-dose evaluation

Abstract

RATIONALE AND OBJECTIVES: Volume perfusion CT (VPCT) of the liver provides functional information but is associated with increased radiation exposure and often requires additional multiphasic CT acquisitions. Photon-counting CT (PCCT) may enable low-dose one-stop imaging by generating diagnostic morphological images directly from perfusion datasets. To evaluate the image quality of single-phase and fused noncontrast and contrast-enhanced images generated from photon-counting CT (PCCT)-based liver dynamic volume perfusion CT (VPCT) for liver morphologic assessment. MATERIALS AND METHODS: In this prospective study, 62 patients with liver disease underwent true noncontrast PCCT, liver VPCT, and low-dose delayed-phase imaging. Single-time-point noncontrast, arterial-phase, and portal venous-phase images were reconstructed from the VPCT dataset using virtual monoenergetic image (VMI, 40-70 keV) and Threshold 3D (T3D) reconstruction. Fused T3D images were generated from multiphase perfusion data. Image quality was evaluated using signal-to-noise ratio (SNR), contrast-to-noise ratio (CNR), subjective image-quality scores. Radiation dose was recorded. RESULTS: The one-stop imaging protocol yielded a lower effective dose than the simulated conventional four-phase protocol combined with perfusion imaging (25.9 ± 6.7 mSv vs 31.9 ± 8.5 mSv, P < 0.05). Compared with true noncontrast images, fused noncontrast images showed higher SNR (8.6 ± 3.0 vs 7.0 ± 1.9) and CNR (1.5 ± 1.2 vs 1.0 ± 0.8) (P < 0.05). Although subjective image-quality scores for perfusion-derived noncontrast images were lower than those for true noncontrast images, all remained diagnostically acceptable (score ≥ 3). In the arterial phase, fused images showed the lowest image noise (7.5 ± 2.7 HU), while 40-keV VMI provided the highest CNR (70.4 ± 24.0). In the portal venous phase, fused images achieved the lowest noise (8.8 ± 2.8 HU) while maintaining relatively high CNR (10.4 ± 7.8). Except for 40-keV and 50-keV portal venous-phase images, all reconstructed arterial- and portal venous-phase images achieved diagnostically acceptable subjective scores (≥3). CONCLUSION: PCCT-based liver VPCT one-stop imaging reduces radiation exposure while providing diagnostically acceptable morphologic images. The optimal reconstruction of noncontrast and contrast-enhanced images varies by imaging phase, supporting a phase-specific image optimization strategy for liver morphologic assessment.