Displayed CTDIvolume accuracy in ultra-low-dose CT: Limitations of current IEC criteria and a proposal for new constancy benchmarks
- Journal
- Radiography (London, England : 1995) (Q2)
- Published
- 10 November 2026
- Study design
- Unclassified
- Evidence level
- Level 5, Expert Opinion (CEBM 5)
- Authors
- T Shohji, R Tomoika, T Kobayashi
- PMID
- 43008525
- DOI
- 10.1016/j.radi.2026.103622
Why clinicians should know about it
- Picked for Medical Physics (top studies of the week, 15 November 2026): Phantom measurements challenge ultra-low-dose CT QC criteria
Abstract
INTRODUCTION: Rapid adoption of ultra-low-dose CT protocols (∼1-2 mGy) used in paediatric cardiac CT examinations challenges current quality control (QC). The IEC 61223-3-5 constancy criterion for CTDIvol display accuracy ("±20% or ±1 mGy") permits up to ±100% relative error at 1.0 mGy, undermining patient dose management and early detection of equipment anomalies. This study evaluates ultra-low-dose display accuracy and proposes revised QC benchmarks. METHODS: CTDIvol accuracy was evaluated across two state-of-the-art CT systems using a cylindrical polymethyl methacrylate (PMMA) phantom with a 32 cm-diameter and a 15 cm-length and a calibrated semiconductor detector. Measurements ranged from 10 to 0.5 mGy (120, 100, and 80 kV), utilizing synchronized and asynchronous rotational exposure methods to ensure geometric reproducibility. RESULTS: Measured doses remained within a ±20% relative tolerance across a tested range from 0.5 mGy to 10.3 mGy, proving precise control is achievable in the ultra-low-dose range. However, system-specific discrepancies were identified, with manufacturers showing distinct overestimation or underestimation tendencies. Crucially, applying the ±1 mGy absolute tolerance below 5.0 mGy permitted exponentially increasing relative errors, effectively nullifying fundamental QC objectives. CONCLUSION: The ±1 mGy absolute constancy criterion loses effectiveness in the ultra-low-dose regime, creating a critical latent risk for undetected equipment degradation. International QC standards urgently require revision to ensure reliable dose monitoring. IMPLICATIONS FOR PRACTICE: We propose establishing a "low-dose boundary value" at 5.0 mGy. Above this threshold, the existing ±1 mGy criterion applies; below it, a uniform ±20% relative tolerance is strictly enforced. This provides a robust, backward-compatible QC framework for low-dose CT, highlighting the need to recognize manufacturer-dependent display uncertainties during protocol optimization.
Abstract as published, via PubMed.
For healthcare professionals. The summary is generated by AI from the published abstract, and the evidence level is assigned automatically from the study design on the Oxford CEBM hierarchy. Neither is medical advice. Read the full paper before changing practice.