Multicentre study of activity quantification methods for 99m-Tc-pyrophosphate SPECT/CT in cardiac amyloidosis

Journal
Physica medica : PM : an international journal devoted to the applications of physics to medicine and biology : official journal of the Italian Association of Biomedical Physics (AIFB) (Q2)
Published
5 November 2026
Study design
Unclassified
Evidence level
Level 5, Expert Opinion (CEBM 5)
Authors
Leonel A Torres, Michael Ljungberg, Anita Brink, Diana Paez, Amalia Peix, Amelia Jimenez-Heffernan, et al.
PMID
42987744
DOI
10.1016/j.ejmp.2026.107217

Why clinicians should know about it

  • Picked for Medical Physics (top studies of the week, 8 November 2026): Multicentre phantom validation of quantitative SPECT activity

Abstract

OBJECTIVE: 99mTc-PYP SPECT/CT imaging is used for diagnosis of transthyretin cardiac amyloidosis (ATTR-CA). The calculation of quantitative metrics allows assessment of the myocardial amyloid burden and supports therapy response evaluations. This study explores the implementation of reliable metrics and the validation of activity quantification methods in a multicentre study. METHODS: As part of the IAEA Transthyretin Amyloid Cardiomyopathy (iTAC) project, a multicentre phantom study was conducted at 10 institutions using 12 hybrid SPECT/CT systems. Standardized acquisitions consistent with clinical protocols were used to compute the activity calibration factors (ACF). Absolute activity verifications were based on cardiac insert measurements with known activity distributions. Multiple reconstruction/quantification approaches (RecMethod) were evaluated, including the institution method (SiteProc) and four alternative ones; results of the computed activity were compared with the true activity meter measurements. Accuracy was assessed using RecMethod-to-True activity ratios, summarized by median, interquartile range (IQR) and coefficient of variation (CV). RESULTS: Most institution ACF values (ACFSiteProc) ranged 75-105 cps/MBq and were consistent with reference CoreLab estimates. Agreement between ACFSiteProc and reference CoreLab values showed a median ratio of 99.6 %, IQR 5.8 %, and CV 9.4 %, indicating no systematic bias. For activity quantification, median RecMethod-to-True ratios were 100.1 % for the SiteProc reconstructions and 97.8 %, 94.1 %, 95.6 % and 96.0 % for the alternatives. CONCLUSION: The evaluated quantification methods demonstrated good median accuracy across institutions providing reliable activity estimates, essential for a robust implementation of quantitative metrics for ATTR-CA. The obtained inter-site variability indicates that precision remains a challenge at the multicentre quantitative imaging studies.

Abstract as published, via PubMed.

View on PubMedFull text at the publisherOpen in the app

For healthcare professionals. The summary is generated by AI from the published abstract, and the evidence level is assigned automatically from the study design on the Oxford CEBM hierarchy. Neither is medical advice. Read the full paper before changing practice.