Skip to main content

Six-year data from three magnetic resonance-guided radiotherapy systems demonstrate long-term stability of linear accelerator quality assurance

In brief

Three MRI-guided radiotherapy systems kept 11 of 13 quality checks within limits

Over six years at one institution, three MRI-guided radiotherapy systems stayed within tolerance limits on 11 of 13 quality checks. The other two checks slightly exceeded limits, but the deviations were understood and not considered clinically relevant; no clinically important worsening trends emerged. The findings support reviewing QA programs, but do not establish which checks can safely be reduced.

Journal
Physics and imaging in radiation oncology (Q1)
Published
23 September 2026
Study design
Cohort / observational study
Evidence level
Level 3, Low (CEBM 3b)
Authors
Guido C Hilgers, Simon J Woodings, Thomas A Foppen, Jochem W H Wolthaus, André W Minken, Bas W Raaymakers, et al.
PMID
42935700
DOI
10.1016/j.phro.2026.101091

Why clinicians should know about it

  • Picked for Medical Physics (paper of the day, 29 October 2026): Six-year measured QA stability across clinical MRI-linacs

Abstract

BACKGROUND AND PURPOSE: Magnetic resonance-guided radiotherapy systems require comprehensive quality assurance (QA) programs to ensure safe and precise treatment delivery. This study evaluates the long-term behavior and variability of linac-related QA results across three clinically operated 1.5 T MRI-linacs at a single institution and assesses compliance with tolerance limits. MATERIALS AND METHODS: QA data acquired over a six-year period from three clinically operated 1.5 T MRI-linacs were retrospectively analyzed. A range of linac-related dosimetric and mechanical QA tests was included. Performance envelopes (mean ± 2 SD) were constructed to characterize variability relative to tolerance limits. RESULTS: For 11 of 13 QA tests, performance envelopes for the associated metrics remained within the applied tolerance limits. For the two remaining tests, envelopes slightly extended beyond these limits. These deviations were not considered clinically relevant, and their underlying causes were understood. No progressive trends of clinical relevance were observed in the linac-specific time series. CONCLUSIONS: Stability of linac-related QA results was demonstrated over six years across three MRI-linacs, supporting reconsideration of the QA program.

Abstract as published, via PubMed.