ESTRO-EPTN guidelines for independent dose calculation and log file-based patient-specific quality assurance in scanned particle therapy

In brief

European guidelines pair independent dose checks with treatment-log review

New European consensus guidelines recommend independent dose calculation to verify the treatment plan and log file review to assess what was delivered in scanned particle therapy. When validated and combined with robust machine quality checks, the methods can detect clinically relevant deviations and reduce reliance on phantom measurements. Their safe use still depends on managing model limits, risks, and data traceability.

Journal
Physics and imaging in radiation oncology (Q1)
Published
10 October 2026
Study design
Practice guideline / consensus
Evidence level
Level 1, High (CEBM 1c)
Authors
Loïc Grevillot, Oliver Jäkel, Abdallah Qubala, Julian Horn, Adam H Aitkenhead, Xiaoying Liang, et al.
PMID
42921085
DOI
10.1016/j.phro.2026.101109

Why clinicians should know about it

Abstract

BACKGROUND AND PURPOSE: Independent dose calculation (IDC) and log file-based patient-specific quality assurance (PSQA) are increasingly used in scanned particle therapy (PT) to address pitfalls of phantom-based verifications. Highly modulated treatment fields, steep dose gradients, and growing interest for adaptive treatments require quality assurance (QA) approaches that are cost effective, patient-specific, and closely linked to the delivered treatment. Practical guidance is therefore needed for the safe clinical implementation of IDC- and log file-based PSQA. MATERIALS AND METHODS: The ESTRO-EPTN (European Particle Therapy Network) developed consensus-based recommendations on IDC- and log file-based PSQA in PT. Current clinical practice, commissioning strategies, risk-management approaches, and available technical solutions were evaluated. Following a review of the literature, expert-based guidelines were developed covering clinical scopes, acceptance, commissioning and validation, integration with machine QA, data management, governance, and staged implementation. RESULTS: IDC- and log file-based PSQA were identified as complementary elements within a quality-management framework. Log file-based approaches enable evaluation of the delivered treatment fraction(s), while IDC provides computational verification of the planned dose. When appropriately validated, these methods increase sensitivity to potentially clinically relevant deviations and can substantially reduce reliance on phantom-based measurements. Safe application requires robust machine QA, characterization of model limitations, risk analysis, and data traceability. Recommendations applicable to all centers were developed for IDC. Log file-based PSQA recommendations include minimum requirements for the vendors. CONCLUSIONS: These recommendations aim to promote consistent and safe adoption across PT centers, while preserving the need for machine QA.

Abstract as published, via PubMed.