Evaluation of EPID-Based Transmission and DLG Correction Methods for Dosimetric Verification in HyperArc Single-Isocenter Multiple Target Radiosurgery
- Journal
- Diagnostics (Basel, Switzerland) (Q2)
- Published
- 19 August 2026
- Study design
- Cohort / observational study
- Evidence level
- Level 3, Low (CEBM 3b)
- Authors
- Se An Oh, Sung Yeop Kim, Jae Won Park, Ji Woon Yea, Jaehyeon Park, Yoon Young Jo
- PMID
- 42651028
- DOI
- 10.3390/diagnostics16162628
Why clinicians should know about it
- Picked for Medical Physics (top studies of the week, 30 August 2026).
Abstract
Background/Objectives: Accurate verification of single-isocenter multiple-target (SIMT) stereotactic radiosurgery is challenging owing to the complexity of multi-lesion delivery and the sensitivity of stereotactic dose gradients. We aimed to evaluate the efficacy of electronic portal imaging device (EPID)-based multileaf collimator (MLC) transmission and dosimetric leaf gap (DLG) correction for improving portal-dose prediction agreement in SIMT stereotactic radiosurgery (SRS) and to propose an exploratory target count-based institutional action level. Methods: This retrospective analysis included 112 consecutive patients treated with HyperArc™-based SRS (1-13 targets). Treatment plans were calculated using the Acuros XB algorithm for 6 MV flattening filter-free (FFF) beams. Two sets of MLC parameters were compared for portal-dose image prediction (PDIP): (1) standard parameters measured using an ion chamber; (2) EPID-derived corrected parameters. Portal-dose accuracy was evaluated using gamma index analysis with a 95% pass rate threshold. Results: The uncorrected method showed a strong negative correlation between target number and gamma passing rates, with complex plans (≥9 targets) dropping as low as 70%. EPID-based correction substantially improved dose agreement, yielding consistent passing rates above 98%, regardless of target number. Although uncorrected parameters remained within tolerance for plans with one to two targets, accuracy declined markedly starting at three targets. Conclusions: Our findings indicate that the EPID-based correction of MLC transmission and DLG mitigated cumulative modeling discrepancies in complex SIMT SRS. Thus, we propose the exploratory institutional observation of three targets, beyond which EPID-based correction may be considered to achieve optimal portal-dose prediction agreement in HyperArc-based stereotactic radiosurgery.
Abstract as published, via PubMed.
For healthcare professionals. The summary is generated by AI from the published abstract, and the evidence level is assigned automatically from the study design on the Oxford CEBM hierarchy. Neither is medical advice. Read the full paper before changing practice.