Time-of-Day-Associated Variation in Serum CO2 Patient-Result Distribution and Its Impact on PBRTQC Performance
In brief
Time-stratified CO2 quality controls cut peak-hour false alarms from 0.955% to 0.069%
In 168,957 results from three analyzers, serum CO2 values ran higher from 10:00 to 13:00, and one set of control limits produced far more false alarms during those hours. Separate time-based limits brought the false-positive rate close to the off-peak level while maintaining error detection; the study's descriptive findings may partly reflect differences in patient case mix.
- Journal
- Diagnostics (Basel, Switzerland) (Q2)
- Published
- 26 October 2026
- Study design
- Cross-sectional study
- Evidence level
- Level 3, Low (CEBM 3b)
- Authors
- Ziqing Jiang, Xuemei Wei, Lei Lin, Jie Feng, Yanhong Gao
- PMID
- 43028377
- DOI
- 10.3390/diagnostics16213504
Why clinicians should know about it
- Picked for Biochemistry (medical) (top studies of the week, 15 November 2026): Directly addresses serum CO2 assay quality and result interpretation
Abstract
Background: Patient-based real-time quality control (PBRTQC) is now widely used in clinical labs, yet the impact of time-dependent variation on its performance remains understudied, particularly with regard to serum carbon dioxide (CO2). This study explores how time-of-day-associated variation in the serum CO2 patient-result distribution affects PBRTQC outcomes and clinical acidosis classification, and an optimization strategy is evaluated. Methods: We analyzed 168,957 serum CO2 results collected over one year (2025) using three biochemical analyzers. Moving average-based PBRTQC was compared between peak (10:00-13:00) and off-peak periods, using either conventional unified limits or time-stratified separate limits. Results: Serum CO2 patient-result distribution shifted significantly toward higher values during the peak period (23.6 vs. 23.0 mmol/L, p < 0.001). A unified control strategy led to a marked rise in false-positive rates (0.955% vs. 0.062%). In a simulated category-transition analysis, 92 of 1212 samples (7.59%) were reclassified, of which 88 were acidosis-to-normal shifts. The time-stratified model reduced peak false positives to 0.069% while maintaining efficient detection of error. Conclusions: Time-of-day-associated variation in the distribution of patients' serum CO2 impairs PBRTQC specificity during peak hours. Time-stratified control limits reduce false-positive rates while maintaining error detection, offering a practical solution for reliable serum CO2 monitoring. These findings should be interpreted as descriptive, given the cross-sectional design and the potential influence of patient case mix.
Abstract as published, via PubMed.