Comparison of automated and manual oxygen titration in emergency care: a systematic review and meta-analysis
In brief
Automated oxygen titration keeps patients in target ranges 23 percentage points longer
In a review of four studies involving 286 adults in emergency care, automated systems increased time within oxygen saturation targets by an average of 23 percentage points compared with manual titration. Results were initially inconsistent, and the studies measured oxygen levels rather than patient outcomes, so whether tighter control improves recovery remains unknown.
- Journal
- CJEM (Q1)
- Published
- 13 November 2026
- Study design
- Systematic review / meta-analysis of RCTs
- Evidence level
- Level 1, High (CEBM 1a)
- Authors
- Laila Choonara, Muhammed Vally, Adam Mahomed, Abdullah E Laher
- PMID
- 43024045
- DOI
- 10.1007/s43678-026-01306-z
Why clinicians should know about it
- Picked for Pulmonary and Respiratory Medicine (top studies of the week, 15 November 2026): Automated oxygen titration improves time within saturation targets
- Picked for Emergency Medicine (top studies of the week, 15 November 2026): Systematic review of oxygen titration in adult emergency care
Abstract
OBJECTIVES: Supplemental oxygen is a cornerstone of emergency care but has a narrow therapeutic window, with both hypoxemia and hyperoxia associated with adverse outcomes. Despite guideline-recommended oxygen saturation targets, manual oxygen titration in acute care settings is frequently inaccurate due to human, system, and workflow limitations. Automated closed-loop oxygen titration systems, which continuously adjust oxygen delivery to maintain predefined saturation targets, have emerged as a potential solution. This systematic review and meta-analysis evaluates the effectiveness of automated oxygen titration compared with standard manual titration in adult emergency care. METHODS: The study protocol was registered with the international prospective register of systematic reviews (PROSPERO). A comprehensive literature search was conducted across PubMed, Scopus, Web of Science, the Cochrane Library, and Google Scholar. Eligible studies involving adult patients receiving emergency care were identified, with full-text review and data extraction performed in accordance with the preferred reporting items for systematic reviews and meta-analyses (PRISMA) guidelines. RESULTS: Twenty-two records were identified; four met inclusion criteria, including three randomized controlled trials (n = 286) and one prospective observational study. In the meta-analysis, automated titration significantly increased time within oxygen saturation target ranges compared to manual care (mean difference 22.9%; 95% CI 9.6-36.2). While initial heterogeneity was high (I2 = 85.7%), sensitivity analysis excluding one high-performing manual cohort eliminated variance (I2 = 0%) and confirmed a robust treatment effect (mean difference 29.6%; 95% CI 23.7-35.6). The available data are limited to these surrogate physiological endpoints; evidence demonstrating improvements in patient-centered clinical outcomes remains absent. CONCLUSION: Automated closed-loop titration demonstrates superior physiological precision in oxygen saturation target maintenance compared to manual care in emergency settings. While it effectively minimizes exposure to hypoxaemia and hyperoxia extremes, further large-scale trials are required to determine if these physiological improvements translate into definitive clinical benefits or improved patient-oriented outcomes.
Abstract as published, via PubMed.
For healthcare professionals. The summary is generated by AI from the published abstract, and the evidence level is assigned automatically from the study design on the Oxford CEBM hierarchy. Neither is medical advice. Read the full paper before changing practice.