Performance of intraoperative artificial intelligence-assisted label-free stimulated Raman histology in recurrent brain tumors

In brief

AI classified recurrent brain tumors correctly in just 39% after multimodal therapy

In this two-center study, AI analysis of rapid tissue images detected tumor with 95.2% sensitivity at an exploratory threshold, while specificity rose to 92%. But classification accuracy fell from 61.8% after surgery alone to 39.0% after surgery, radiation, and medical treatment, and nearly a third of images lacked sufficient quality. Reliable use will depend on better image-quality controls and models adapted to treated tumors.

Journal
Neuro-oncology advances (Q1)
Published
29 October 2026
Study design
Cohort / observational study
Evidence level
Level 3, Low (CEBM 3b)
Authors
Gina Fürtjes, Abdulkader Al Shughri, Thomas Stehle, David Reinecke, Anna-Katharina Meißner, Nina Jung, et al.
PMID
42992275
DOI
10.1093/noajnl/vdag267

Why clinicians should know about it

  • Picked for Pathology and Forensic Medicine (paper of the day, 8 November 2026): AI-assisted intraoperative histology evaluates tumour detection and classification

Abstract

BACKGROUND: Therapy-induced tissue alterations complicate intraoperative diagnosis of recurrent central nervous system (CNS) tumors. We evaluated stimulated Raman histology (SRH) combined with convolutional neural networks (CNNs) for rapid tumor detection and classification. METHODS: In this bicentric observational study, fresh tissue from patients undergoing resection or stereotactic biopsy for suspected recurrent CNS tumors was imaged with fiber-laser-based SRH. CNN1 performed binary tumor/non-tumor prediction and CNN2 histotype classification. CNN1 was evaluated against blinded neuropathologist SRH consensus and CNN2 against the final integrated histopathological diagnosis. RESULTS: Of 382 acquired SRH images, 262 images from 143 patients were included. CNN1 achieved an AUC of 0.975 (95% CI, 0.95-1.00) for tumor detection. An exploratory cohort-derived tumor probability threshold of ≥59 increased specificity from 70.7% to 92.0% while maintaining 95.2% sensitivity and identifying 71.4% of radiation necrosis samples. Among 212 images in treatment subgroup analysis, CNN1 accuracy did not differ significantly across prior-treatment groups (overall 90.1%; P = .080). CNN2 accuracy declined from 61.8% after surgery alone to 39.0% after surgery, radiotherapy, and medical treatment (χ2 = 6.04, P = .049). Multimodal therapy was independently associated with lower odds of correct CNN2 prediction (adjusted OR, 0.10 [95% CI, 0.02-0.39]; P = .001 and 0.04 [95% CI, 0.01-0.20]; P < .001). 120/382 images (31.4%) were of insufficient diagnostic quality; application of the quality control threshold would have rejected 84/120 (70.0%). CONCLUSIONS: CNN1 maintained high tumor/non-tumor discrimination across treatment subgroups when image quality was sufficient, whereas CNN2 entity-classification performance declined with increasing treatment intensity. Reliable SRH-based diagnosis in recurrent tumors requires validated image-quality control measures and model adaptation to recurrent disease.

Abstract as published, via PubMed.