Biomechanical effects of lateral pedicle wall violation and subsequent screw redirection on pedicle screw anchorage

In brief

In vertebra tests, redirecting breached screws cut anchorage cycles another 33%

In tests of 20 human thoracolumbar vertebrae, screws that breached the pedicle wall endured 36% fewer load cycles before loosening than standard screws. Redirecting a breached screw into a standard path with the same dimensions reduced endurance by a further 33% compared with leaving it in place. These laboratory results challenge routine redirection for better fixation, but do not establish clinical outcomes.

Journal
European spine journal : official publication of the European Spine Society, the European Spinal Deformity Society, and the European Section of the Cervical Spine Research Society (Q1)
Published
28 August 2026
Study design
Randomized controlled trial
Evidence level
Level 1, High (CEBM 1b)
Authors
Anna Spicher, Richard A Lindtner, Tim Egelseer, Rene Schmid, Peter W Ferlic, Romed Hörmann, et al.
PMID
42663627
DOI
10.1007/s00586-026-10333-4

Why clinicians should know about it

  • Picked for Spine Surgery (top studies of the week, 18 October 2026): LPWV reduces pedicle screw anchorage

Abstract

PURPOSE: To investigate the biomechanical effects of lateral pedicle wall violation (LPWV) and subsequent screw redirection on pedicle screw anchorage under cyclic loading. METHODS: Twenty fresh-frozen human thoracolumbar vertebrae (T10-L1) were randomly allocated to two experimental groups (n = 10 each). In Group 1, each vertebra received one standard intrapedicular screw and one LPWV screw, with screw trajectories randomly assigned to the left or right pedicle. In Group 2, both pedicles were initially instrumented with LPWV screws; one screw per vertebra was randomly selected for redirection into a standard intrapedicular trajectory, while the contralateral LPWV screw was left in place. All screws were subjected to cyclic cranio-caudal loading with stepwise increasing compressive load until loosening. RESULTS: LPWV screws withstood significantly fewer load cycles until loosening than standard intrapedicular screws (5230 ± 2481 vs. 8205 ± 3027 cycles; p = 0.003), corresponding to a 36% reduction in anchorage. Redirected screws withstood significantly fewer load cycles until loosening than LPWV screws left in place (3686 ± 2105 vs. 5489 ± 2731 cycles; p = 0.018), representing an additional 33% reduction. CONCLUSION: LPWV substantially compromises pedicle screw anchorage under cyclic loading, even when the screw remains at least half contained within the pedicle and fully contained within the vertebral body. Redirection of LPWV screws using the same diameter does not restore fixation strength but is associated with a further reduction in anchorage, challenging the assumption that redirection inherently improves screw stability. Redirection of LPWV screws using screws of the same dimensions and design may therefore be considered with caution, particularly in the absence of neural or vascular compromise.

Abstract as published, via PubMed.

View on PubMedFull text at the publisherOpen in the app